Spectrometric electronics
2015.08.06 14:18 - adminWe offer a set of modular electronic instruments that together with Tukan8k multichannel analyser make up a complete spectrometer of ionising radiation. The set consists of a NaI(Tl) scintillation detector, a HV power supply, and a spectrometric pulse amplifier. Each module is implemented on a 3U x 220 mm printed circuit board. All modules are housed together in a standard EURO 3U cassette.
NaI(Tl) scintillation detector
Dimensions of the NaI(Tl) crystal | 2"x2" |
Photomultiplier | XP3212B |
Energy resolution | ≤ 7% for Cs137 |
High voltage divider, integrating preamplifier, and magnetic shield integrated within the detector housing. |
1.5 kV power supply
Output | regulated 0…1.5 kV, + polarity |
Rated output current | 2 mA |
Output temperature stability | ±0.01%/oC |
Output load stability | ±0.001% |
Ripples | less than 0.008% p-p |
Voltage reading | 4-digit LED display |
Spectrometric pulse amplifier
Gain (coarse and fine regulation) | 2.5…1500 V/V |
Shaping time constant | 0.5–1–2–4 µs |
Integral non-linearity | better than ±0.05% |
Noise (at 2 μs shaping, gain >100 V/V) | ≤ 8 µV RMS related to input |
Polarity of input pulses | +/–, tailing time constant ≥ 40 μs |
Output pulses | unipolar/bipolar, linear up to 10V |
Additional features: |
Pole-zero compensation circuit (P/Z) Gated base-line restorer circuit |
Contact Eng. Dariusz Wolski (phone 048 22 273 1547, e-mail dariusz.wolski@ncbj.gov.pl) to get more technical information.